| Model Number |
20226 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that stent dislodgement occurred, requiring an additional device.The target lesion was located in the mildly tortuous and non-calcified iliac and renal arteries.Following pre-dilatation with a 6.0 x 80, 135cm mustang balloon catheter and placing a new 7f non-boston scientific sheath, a 8.0x40x75cm express ld stent was selected for use.However, when the express ld went over the bifurcation and was just about to be placed over the lesion, the stent dislodged.The physician decided to leave the stent and deployed it using a new mustang balloon catheter.Another express ld stent was used to cover the remainder of the lesion.There were no complications reported and the patient was scheduled for follow-up.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the express ld was returned for analysis.The balloon was received loosely folded.The balloon appears to have been subjected to positive pressure.No stent received on balloon material.A visual and tactile examination identified no kinks or damage to the shaft of the device.A visual examination of the device identified no issues with the markerbands or tip.
|
| |
|
Event or Problem Description
|
|
It was reported that stent dislodgement occurred, requiring an additional device.The target lesion was located in the mildly tortuous and non-calcified iliac and renal arteries.Following pre-dilatation with a 6.0 x 80, 135cm mustang balloon catheter and placing a new 7f non-boston scientific sheath, a 8.0x40x75cm express ld stent was selected for use.However, when the express ld went over the bifurcation and was just about to be placed over the lesion, the stent dislodged.The physician decided to leave the stent and deployed it using a new mustang balloon catheter.Another express ld stent was used to cover the remainder of the lesion.There were no complications reported and the patient was scheduled for follow-up.
|
| |
|
Search Alerts/Recalls
|