• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE A-OK OPHTHALMIC KNIVES; KNIFE, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE A-OK OPHTHALMIC KNIVES; KNIFE, OPHTHALMIC Back to Search Results
Catalog Number 8065921501
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Codes Corneal Edema (1791); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that during surgery, an ophthalmic blades were noted to be dull.The patient experienced corneal edema, descemet was completely missing and it was treated with topical anti inflammatory drops.The current patient status was not provided.
 
Additional Manufacturer Narrative
Additional information was provided in sections h.6 and h.11.A sample was not received at the manufacturing site for evaluation for the report of dull blades /corneal edema /descement missing /additional intervention; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
A-OK OPHTHALMIC KNIVES
Common Device Name
KNIFE, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18845691
Report Number2523835-2024-00257
Device Sequence Number12845328
Product Code HNN
UDI-Device Identifier00380659215018
UDI-Public00380659215018
Combination Product (Y/N)N
Initial Reporter StateWY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065921501
Device Lot Number168UEN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2024
Supplement Date Received by Manufacturer05/10/2024
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date06/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-