• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTRA-LOCK INTERNATIONAL, INC. INTRA-LOCK INTERNATIONAL, INC.; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTRA-LOCK INTERNATIONAL, INC. INTRA-LOCK INTERNATIONAL, INC.; DENTAL IMPLANT Back to Search Results
Model Number SCT4711
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/08/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Cause of the issue: during the migration of biohorizon¿s account from development to production, the fda gateway was mapped to emdr form version 2015 instead of 2021.The incorrect mapping to form version 2015 resulted in the submissions failing to move through the fda gateway.Impact: the records impacted were stuck in "submission in progress" state causing them to not advance to the fda.Impacted records count: 480 emdr records out of a total 1238 emdr's in the system.Biohorizons: biohorizons is issuing a supplier corrective action request (scar) to our vendor sparta.This will allow biohorizons to formally document the issue, actions taken by sparta to correct, and along with sparta actions taken to prevent the reoccurrence of this event.Scar number scar24001 was issues to sparta to document the event.Sparta: to resolve the issue, sparta corrected the emdr form version mapping to 2021 for the biohorizons production account.The fix does not retrospectively re-submit the impacted records to the fda.To accomplish this additional step are needed to re-submit the impacted records to the fda.These steps are: 1.Update the mdr submission status on each record to a blank value.2.Update h10 field with mfg narrative for the late submission.3.Using trackwise digital platform automation, progress the impacted records to "failed fda." 4.Using trackwise digital platform automation, progress the impacted records to "submission in progress fda." preventative actions: this sequence of events which led to the issue will not occur again for biohorizons because the emdr form version mapping is a one-time activity.During the time of production deployment, biohorizons was a new customer of sparta.Sparta has raised the event to the product organization in exploring the options to remove any retired fda form references from the system.One of the options is for new customers to remove the old form version and verify the current fda version is appearing within the new customers production environment.Moreover, to facilitate early identification of fda submission issues, a report will be built to identify emdr records in "submission in progress" state for a defined period of time.
 
Event or Problem Description
Implant failed to osseointegrate.
 
Event or Problem Description
Implant failed to osseointegrate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRA-LOCK INTERNATIONAL, INC.
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
INTRA-LOCK INTERNATIONAL, INC.
2300 riverchase center
birmingham AL 35244
Manufacturer (Section G)
INTRA-LOCK INTERNATIONAL, INC.
2300 riverchase center
birmingham AL 35244
Manufacturer Contact
anissa smith
2300 riverchase center
birmingham, AL 35244
MDR Report Key18846208
Report Number3003631996-2024-00006
Device Sequence Number15392891
Product Code DZE
UDI-Device IdentifierD742SCT4711
UDI-PublicD742SCT4711
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K103194
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/18/2026
Device Model NumberSCT4711
Device Catalogue NumberSCT4711
Device Lot Number2101109
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2024
Supplement Date Received by Manufacturer01/19/2024
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-