• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949 SYRINGE 0.5ML 31GA 8MM 10 BAG 500 WAL; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949 SYRINGE 0.5ML 31GA 8MM 10 BAG 500 WAL; SYRINGE, PISTON Back to Search Results
Model Number 328509
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported she was not able to draw insulin.Stated, she saw a "air bubble" inside the barrel when trying to draw.Lot: 3156431, catalog: 328509, date of event: unknown, samples: yes cl.
 
Additional Manufacturer Narrative
Product investigation is in progress.Once it is completed, the investigation summary will be submitted.
 
Additional Manufacturer Narrative
Additional information was added to: b4, g6, h2, h3, h11.Correction to: h6 (investigation findings, and investigation conclusions).Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.5ML 31GA 8MM 10 BAG 500 WAL
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949
Manufacturer (Section G)
BECTON DICKINSON AND CO.
1 becton drive
franklin lakes, NJ 07417
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key18847295
Report Number1920898-2024-05053
Device Sequence Number7814279
Product Code FMF
UDI-Device Identifier00681131311762
UDI-Public00681131311762
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number328509
Device Lot Number3156431
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2024
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date06/01/2024
08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-