| Model Number |
775100 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported they are receiving multiple errors such as downstream occlusion errors, the tubes are leaking water, and air is getting into the tubing.Per additional information provided on (b)(6) 2024, the customer stated that from the nursing reports it sounds like it was either leaking prior to priming or after priming, the tubes got air in them, and that caused the error messages.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction made to d4.Model and catalog numbers corrected to 775100.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on april 27, 2023.One decontaminated sample was received at the manufacturing site for evaluation.Visual and functional evaluations were performed, and the reported condition could not be confirmed.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be used for tracking and trending purposes.
|
| |
|