• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS PHOS2; PHOSPHOMOLYBDATE (COLORIMETRIC), INORGANIC PHOSPHORUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS PHOS2; PHOSPHOMOLYBDATE (COLORIMETRIC), INORGANIC PHOSPHORUS Back to Search Results
Catalog Number 05171377190
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with phos2 (phosphorus) on a cobas 8000 c702 module.The sample initially resulted in a phosphorus value of 0.9 mg/dl.Due to the value being low, the sample was repeated, resulting in a value of 9.5 mg/dl.The sample was repeated several times, but not additional values were provided.The 0.9 mg/dl value was deemed correct.
 
Additional Manufacturer Narrative
The serial number of the c702 analyzer is (b)(6).The investigation is ongoing.
 
Additional Manufacturer Narrative
Calibration and controls were acceptable.The customer indicated they did not receive any alarms.The field service engineer could not determine a cause.No issues were observed with reaction cell washing, probe washing, or probe adjustments.No issues were observed with calibration or controls.The investigation could not identify a product problem.The cause of the event could not be determined.The customer has not reported any additional issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHOS2
Common Device Name
PHOSPHOMOLYBDATE (COLORIMETRIC), INORGANIC PHOSPHORUS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18847427
Report Number1823260-2024-00670
Device Sequence Number15407411
Product Code CEO
UDI-Device Identifier04015630925094
UDI-Public04015630925094
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K951595
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05171377190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/20/2024
Supplement Date Received by Manufacturer04/05/2024
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date04/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACETAMINOPHEN
Patient Age86 YR
Patient SexMale
Patient Weight66 KG
-
-