| Catalog Number |
05171377190 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with phos2 (phosphorus) on a cobas 8000 c702 module.The sample initially resulted in a phosphorus value of 0.9 mg/dl.Due to the value being low, the sample was repeated, resulting in a value of 9.5 mg/dl.The sample was repeated several times, but not additional values were provided.The 0.9 mg/dl value was deemed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The serial number of the c702 analyzer is (b)(6).The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration and controls were acceptable.The customer indicated they did not receive any alarms.The field service engineer could not determine a cause.No issues were observed with reaction cell washing, probe washing, or probe adjustments.No issues were observed with calibration or controls.The investigation could not identify a product problem.The cause of the event could not be determined.The customer has not reported any additional issues.
|
| |
|