| Catalog Number |
08946370190 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cobas e801 analytical unit serial number was (b)(6).The customer suspected a sample interference.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 1 patient's sample tested with elecsys testosterone g2 (testosterone ii) assay on a cobas e801 immunoassay analyzer when compared to the liquid chromatography mass spectrometry (lcms) method.The initial result was reportedly 1192 ng/dl.On (b)(6) 2023 a new sample was drawn from the patient and the result was reportedly 899 ng/dl.On (b)(6) 2023 a new sample was drawn from the patient and the result was reportedly 854 ng/dl.The provider questioned the results as they did not match the patient's history.On (b)(6) 2023 a new sample was drawn from the patient and tested with lcms method.The result was reportedly approximately 200 ng/dl.The exact result was not provided.The result tested with lcms method was deemed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The calibration was last performed on 15-feb-2024 and was acceptable.Data for investigation was requested but not provided.The customer declined the service visit.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|