DEPUY MITEK LLC US VAPR 2.3MM WEDGE 1-PIECE ELECTRODE -EA; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
227203 |
| Medical Device Problem Codes |
Break (1069); Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi:(b)(4).Investigation summary: according to the information received, it was reported that during the surgery, when in use, noted many bubbles (as the video shows), and is was found the ceramic was broken (as the photo shows).Another device was used to complete the surgery.There were no adverse consequences to the patient.The photos and video investigation revealed that vapr 2.3mm wedge 1-piece electrode -ea was producing bubbles when activated.The ceramic insulator is cracked.The overall complaint was confirmed as the observed condition of the vapr 2.3mm wedge 1-piece electrode -ea would contribute to the complained device issue.A manufacturing record evaluation was performed for the finished device lot number (u2302084) and no non-conformances were identified.Based on the investigation findings, the potential cause for the cracked ceramic insulator can be attributed to unintended use error by use of excessive force in a prying motion and overload of the material.The potential cause for the bubbles will be determined once the device is evaluated physically.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Event or Problem Description
|
|
It was reported by the healthcare professional in china that during anterior/posterior repair procedure on (b)(6) 2024, when using the vapr 2.3mm wedge 1-piece electrode device many bubbles were observed and it was found that the ceramic was broken.Another like device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The investigation has been updated to reflect the correct information: investigation summary: the photos and video investigation revealed that vapr 2.3mm wedge 1-piece electrode -ea was producing bubbles when activated.The ceramic insulator shows a line that appears to be a crack.The manufacturer performed the video evaluation with the following result: technical authority have reviewed the video and observed the complaint device through additional testing.The return is surrounded by tissue, this acts as a high impedance, so when the return area is reduced to less than the active tip it then becomes the active this is why there are bubbles coming from the return tube, it¿s now acting as the active tip.There are no breakages or leaks around the ceramic there is a warning in the product ifu to make sure the electrodes are surrounded by conductive irrigant during use.Therefore the conclusion of ¿no fault found¿ is correct.The overall complaint was confirmed as the observed condition of the vapr 2.3mm wedge 1-piece electrode -ea would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Investigation summary: the device was received and evaluated.Visual inspection revealed that the electrode is in used condition.The cable is in good condition as well as the connector and pins.The active tip shows signs of activation.The ceramic tip shows a line that appears to be a crack.When connected to the test generator, the electrode was immediately recognized.When activating the electrode, very small bubbles were coming out of the tip.The device will be sent to the manufacturer for further analisys.Manufacturer evaluation result for vapr 2.3mm wedge 1-piece electrode -ea: the device was received in shelf carton.Device was in used condition.No visible crack on ceramic only surface marks.From the ceramic coming in to contact with another object.Handle assembly was in good condition.Cable and plug were in good condition.Electrical test: the device passed all the parameters.Functional test: the device passed all the checks.From our investigation we were unable to confirm either of the customer reported issue(s) visual inspection showed only surface marks from the ceramic coming in to contact with another object, no cracks or structural anomalies were observed.Activation of the electrode did not reveal any anomalies with the degree of bubble creation being as expected.Testing at atl uk showed the returned device was immediately recognized and found to pass both electrical and functional testing, with no abnormal bubble creation or any other issue being detected.Activation was found to be consistent and as expected.The returned complaint device was found to meet the manufacturers specification; therefore, no further investigation is possible regarding the returned complaint device.Conclusion: no fault found - the root cause of the customer reported issues(s) could not be determined.Technical authority have reviewed the video and observed the complaint device through additional testing.The return is surrounded by tissue, this acts as a high impedance, so when the return area is reduced to less than the active tip it then becomes the active this is why there are bubbles coming from the return tube, it¿s now acting as the active tip.There are no breakages or leaks around the ceramic there is a warning in the product ifu to make sure the electrodes are surrounded by conductive irrigant during use.Therefore the conclusion of ¿no fault found¿ is correct.The overall complaint was unconfirmed as the observed condition of the vapr 2.3mm wedge 1-piece electrode -ea would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review : a manufacturing record evaluation was performed for the finished device u2302084 number, and no non-conformances were identified.
|
| |
|