• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL EZ-IO 45MM NEEDLE (BOX OF 5); NEEDLE, HYPODERMIC, SINGLE LU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL EZ-IO 45MM NEEDLE (BOX OF 5); NEEDLE, HYPODERMIC, SINGLE LU Back to Search Results
Catalog Number 9079-VC-005
Medical Device Problem Codes Crack (1135); Leak/Splash (1354)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 01/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that: malfunction - during or after use.Additional information: propofol and fentanyl were infusing through io when it started leaking.User attempted to undo the extension set and reattach but continued to leak, then a 10cc syringe was attached to the io while they went to get a new extension set but leak continued through a crack in the io.The user felt it may be marrow leaking.The patient did have a central line which had medication running at the time and they attempted to insert a peripheral line but failed.Medication was swapped out of the central line for the sedation and the io was removed.Impact: the patient was not sedated when they should have been which caused discomfort.
 
Additional Manufacturer Narrative
Qn#(b)(4).The complaint cannot be confirmed.Complaint verification testing could not be performed as no sample was returned for analysis.A review of the certificate of compliance was performed with no relevant findings.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that: malfunction - during or after use.Additional information: propofol and fentanyl were infusing through io when it started leaking.User attempted to undo the extension set and reattach but continued to leak, then a 10cc syringe was attached to the io while they went to get a new extension set but leak continued through a crack in the io.The user felt it may be marrow leaking.The patient did have a central line which had medication running at the time and they attempted to insert a peripheral line but failed.Medication was swapped out of the central line for the sedation and the io was removed.Impact: the patient was not sedated when they should have been which caused discomfort.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ-IO 45MM NEEDLE (BOX OF 5)
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LU
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18847860
Report Number3011137372-2024-00026
Device Sequence Number7832261
Product Code FMI
UDI-Device Identifier00816000011980
UDI-Public00816000011980
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K180395
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9079-VC-005
Device Lot Number69554796
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2024
Supplement Date Received by Manufacturer03/27/2024
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date05/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Patient SexUnknown
-
-