| Model Number |
29711 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the wire detached, and additional surgery was required.The 90% stenosed target lesion was located in the severely calcified artery.A rotawire was selected for atherectomy.During procedure, following rotablation, a sterling balloon would not advance over the wire and the wire appeared kinked.The wire then broke into two pieces, mid wire.Furthermore, the patient underwent endarterectomy and the distal end of the rotawire was successfully removed.The procedure was completed.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: lot number: corrected from unknown to (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the wire detached, and additional surgery was required.The 90% stenosed target lesion was located in the severely calcified artery.A rotawire was selected for atherectomy.During procedure, following rotablation, a sterling balloon would not advance over the wire and the wire appeared kinked.The wire then broke into two pieces, mid wire.Furthermore, the patient underwent endarterectomy and the distal end of the rotawire was successfully removed.The procedure was completed.No patient complications were reported.
|
| |
|