DEPUY MITEK LLC US RIGIDFIX CURVE GUIDE FRAME; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number 213004
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: e3: reporter is a j&j sales representative.H4: the device manufacture date is unknown.Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
It was reported by the sales rep that during an unspecified surgical procedure on an unknown date the rigidfix curve guide frame device had a broken screw in the curve set.It was impossible to use the guide without that screw.In order to resolve the issue, the surgeon used a back-up set.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4: the device date of manufacture was unknown in the initial report and has been updated accordingly.Investigation summary: photos were returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection of the returned photos.Upon visual inspection of the photos, it was noted that the device screw head is broken, also the device shows marks of wear, no further information was provided.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the rigidfix curve guide frame would contribute to the complained device issue.Based on the investigation findings, the root cause is traced to end of life.As per the instructions for use; inspect all instruments for damage before use.Do not attempt to repair.It has been determined, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance device history review : a manufacturing record evaluation was performed for the finished device lot number: and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGIDFIX CURVE GUIDE FRAME
Common Device Name
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key18849552
Report Number1221934-2024-00747
Device Sequence Number15260567
Product Code HTY
UDI-Device Identifier10886705019953
UDI-Public10886705019953
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K130105
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number213004
Device Lot Number136725RR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date04/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2014
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-