| Model Number |
PCF-Q180AI |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the colonovideoscope c-cover (plastic distal end cover) had foreign objects.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and the evaluation found no malfunctions aside from the allegation reported in b5.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigations, the adhesion/clogging of the material at c-cover was confirmed.It could not be specified what the foreign material was nor the root cause of the remaining foreign material.The foreign material was possibly not removed since the reprocessing was not conducted properly due to leakage from the remote switch 1.Performance of reprocessing on the device was secured in the reports by reprocessing appropriately in accordance with instructions for use (ifu/cleaning/disinfection/sterilization).There was no report that the device was used for procedure while the event occurred.The suggested events can be detected and prevented by handling the device in accordance with the following instructions for use (ifu): ifu states that detection method in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection.Ifu states that preventive measure in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).A definitive root cause cannot be determined.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|