Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Medtronic received information 11 years post implant of this 25mm aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported as severe aortic regurgitation (ar) with the aortic valve peak velocity is 2.7 m/s.The velocity ratio is 0.37.The mean gradient is 16 mmhg.The peak gradient is 28 mmhg.The aortic valve area by continuity equation is 1.2 cm2.No intervention or adverse patient effects were reported.Subsequently 4 days later, the patient had an aortic valve-in-valve replacement with a transcatheter valve.No additional adverse patient effects were reported.
|