| Model Number |
72404302 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this inflatable penile prosthesis (ipp) is no longer functioning as expected.An ultrasound was performed, and it was found that there was no fluid left in the reservoir which indicates fluid leakage at an unspecified location.The device remains implanted, and no patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available, the cause that contributed to the reported event cannot be established; a conclusion code of cause not established was assigned to this investigation.
|
| |
|