• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3341359
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture, capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient implanted with a 445cc (left), 495cc (right), mentor memoryshape breast implant experienced bilateral ruptures, capsular contracture (baker grade 4) post-operatively.As a result, the patient underwent explantation on (b)(6) 2024.This report is for the right side device.
 
Additional Manufacturer Narrative
On march 7, 2024, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation. mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that the mentor memoryshape¿ tm+ 495cc breast implant was found to have some bubbles within the gel.Leak testing was performed, in accordance with mentor procedures, and no areas of gel exposure were detected during the analysis.Bubbles in the gel can originate with changes in pressure and temperature which can affect volume.The shell wall is permeable to air, therefore trapped air can form bubbles with changes in pressure or temperature.When left by themselves with no changes in temperature or pressure, bubbles usually dissipate over time, unless trapped by cured gel.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture status identified by mri evaluation includes both suspected ruptures, which are those ruptures identified by mri but not confirmed by explantation and examination of the device, and confirmed ruptures, which are those ruptures that are confirmed by evaluation of the explanted devices.The event described could not be confirmed as the breast implant was returned without detectable gel exposure.Although no rupture was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Mentor received a confirmed date of implant.Relevant fields have been updated.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key18851876
Report Number1645337-2024-02800
Device Sequence Number7485654
Product Code FTR
UDI-Device Identifier00081317005100
UDI-Public00081317005100
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/25/2021
Device Catalogue Number3341359
Device Lot Number7310264
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2024
Supplement Date Received by Manufacturer03/07/2024
04/23/2024
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date03/18/2024
04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/29/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-