MEDOS INTERNATIONAL SARL RGDLOOP ADJUSTABLE LONG; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 232448
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Udi: (b)(4).Investigation summary: the product has not returned to depuy synthes mitek, however a photo was provided for review.The photo investigation revealed that rgdloop adjustable long had the white suture frayed, the green suture had foreign matter, presumably biological matter.No other anomalies could be observed as the suture was tangled.The overall complaint was confirmed as the observed condition of the rgdloop adjustable long would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to procedures variables such as: snaps or other instrument to pull the suture without wrapping the suture around the snaps creates a stress point on the suture, resulting in breaking it.The white suture is used to assembly into a graft construction with provides fixation.The green / white suture is the lead suture due to it is used to pull the implant/ graft construction with tension, the tension values are associated with this final implant assembly; the strength requirement adjusting suture loops from dve testing averaged well over the 250n clinical spec ~1700n.The tension clinical spec is high compared to expected intraoperative loads is 20-50n.Therefore, based on the work instruction of finish goods, the tension is measured only on the green/white suture during the manufacturing process.Regarding the white suture, an inspection is performed during the manufacturing process to check the measurement value and the condition; this information correspond to a process control and it is the document to use to guarantee the inspection of the suture, due to this damage suture can¿t have happened during manufacturing process.As per instructions for use (ifu), the steps for a proper insertion are provided, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event or Problem Description
It was reported by the sales rep in china that during a cruciate ligament reconstruction surgical procedure performed on (b)(6) 2024 it was observed that when the rgdloop adjustable long suture device was tightened, the suture was broken off.It was reported that another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the product was returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection of the returned device.Visual inspection found that rgdloop adjustable long white suture was frayed and broken.Also, foreign matter was on the green suture, presumably biological matter.The overall complaint was confirmed as the observed condition of the rgdloop adjustable long would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to the procedural variables, such handling of the device or product interaction during procedure such as: snaps or other instrument to pull the suture without wrapping the suture around the snaps creates a stress point on the suture, resulting in breaking it.The white suture is used to assembly into a graft construction with provides fixation.The green / white suture is the lead suture due to it is used to pull the implant/ graft construction with tension, the tension values are associated with this final implant assembly; the strength requirement adjusting suture loops from dve testing averaged well over the 250n clinical spec ~1700n.The tension clinical spec is high compared to expected intraoperative loads is 20-50n.Therefore, based on the work instruction of finish goods, the tension is measured only on the green/white suture during the manufacturing process.Regarding the white suture, an inspection is performed during the manufacturing process to check the measurement value and the condition; this information correspond to a process control and it is the document to use to guarantee the inspection of the suture, due to this damage suture can¿t have happened during manufacturing process.As per instructions for use (ifu), the steps for a proper insertion are provided, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RGDLOOP ADJUSTABLE LONG
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT MITEK
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key18852004
Report Number1221934-2024-00767
Device Sequence Number7486154
Product Code MBI
UDI-Device Identifier10886705024100
UDI-Public10886705024100
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K140324
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue Number232448
Device Lot Number194L224
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2024
Supplement Date Received by Manufacturer03/19/2024
03/20/2024
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date03/20/2024
03/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-