• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-42-13
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Dislocation (2374)
Date of Event 01/17/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(d10) concomitant device(s): 320-10-10 - equinoxe reverse tray adapter plate tray +10: 5232147.320-02-42 - rs expanded glenosphere 42mm, +4mm offset: 5645362.320-15-05 - eq rev locking screw: 5600768.320-20-00 - eq reverse torque defining screw kit: 5667604.(h3) pending evaluation.
 
Event or Problem Description
As reported by the equinoxe shoulder study, the patient had a left tsa on (b)(6) 2018 and presented with disassociation of polyethylene, on (b)(6) 2024.Patient dislocated his reverse and disassociated his poly component getting off the floor.The outcome of this event is considered resolved on 1/17/2024.The action taken was revision on (b)(6) 2024 with the removal of the torque screw, humeral tray, humeral liner, glenosphere, glenosphere locking screw.The case report form indicates that this event is definitely not related to the device and/or to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported may have been due to disassembly and dislocation.However, the reported disassembly and dislocation could not be confirmed.Potential contributions of user and patient-related considerations to the event, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
miguel sosa
2320 nw 66th ct.
gainesville, FL 32653
3523771140
MDR Report Key18852012
Report Number1038671-2024-00413
Device Sequence Number7486158
Product Code KWT
UDI-Device Identifier10885862086723
UDI-Public10885862086723
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/25/2022
Device Catalogue Number320-42-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2024
Supplement Date Received by Manufacturer05/06/2025
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/27/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient SexUnknown
-
-