Blank fields on this form indicate the information is unknown, unchanged or unavailable.Investigation ¿ evaluation it was reported that on 19feb2024, the gauze in a cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray had a fly in it.The device did not make patient contact.The procedure was completed by using a different gauze package.The patient did not experience any adverse effects or require any additional procedures due to this event.Reviews of the documentation, including the complaint history, device history record, quality control procedures, instructions for use, as well as a visual inspection of the returned device, were conducted during the investigation.Cook received a few pieces of gauze from the customer.Visual inspection of the returned device confirmed that an insect was present.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for the reported complaint device lot revealed no relevant nonconformances that could have contributed to the reported failure mode.It should be noted that there were no other complaints associated with the final product lot number.Cook also reviewed product labeling, the product ifu, [c_t_ulmabrm_rev4] ¿spectrum and spectrum glide central venous catheters,¿ provides the following information to the user related to the reported failure mode: how supplied: ¿do not use the product if there is doubt as to whether the product is sterile.¿ evidence gathered upon review of the dmr, dhr, product labelling and returned device suggests that the device was manufactured out of specification.There is no evidence of additional nonconforming material in house or in the field.Based on the information provided, examination of the returned product, and the results of our investigation, it was concluded that the cause of this event is a quality control deficiency.Per the risk assessment no further action is required.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|