• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 3 CEM; KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 3 CEM; KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150680003
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Joint Laxity (4526); Unspecified Tissue Injury (4559)
Date of Event 02/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
All components were removed except for the patella.Reason for revision was due to instability, particularly mid flexion instability.They have had both sides done, but were never happy with this side.Components were all well fixed.Surgeon noted that the tibial slope did not look right, and there was also bone left out the back from the primary tibial resection which was not adequately cleared.They were also tighter in extension relative to flexion, so was not a balanced knee.Doi: (b)(6) 2021 doe: (b)(6) 2024.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, "[reason for revision was due to instability, particularly mid flexion instability.Surgeon noted that the tibial slope did not look right, and there was also bone left out the back from the primary tibial resection which was not adequately cleared.They were also tighter in extension relative to flexion, so was not a balanced knee.Revision was as a result of a poorly done primary, not implant failure.]" the product was not returned to depuy synthes, however photos were provided for review.See attachment (screenshot (26)).The photographs attached were reviewed, however they do not represent the reported complaint condition.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the photographs provided are not representative of the reported complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIB BASE SZ 3 CEM
Common Device Name
KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18852251
Report Number1818910-2024-05237
Device Sequence Number7486213
Product Code NJL
UDI-Device Identifier10603295492153
UDI-Public10603295492153
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number150680003
Device Lot Number9660331
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/18/2024
Initial Report FDA Received Date03/06/2024
Supplement Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE MEDIAL ANAT PAT 29MM; ATTUNE PS FEM LT SZ 4 CEM; ATTUNE PS RP INSRT SZ4 5MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-