The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the segment fluid damage, the ccd module with drive pcb fluid damage, the lcb (light carrying bundle) fluid damage, the rubber trim collar broken, the lcb (light carrying bundle) broken, the insertion flexible tube buckled, the light guide cable buckled, the root brace rubber (insertion flexible tube) cracked, the control body corroded, the lg connector corroded, the operation channel (primary) leak, and the operation channel (primary) resistance; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|