| Model Number |
SUTURE ANCHOR, BIO-CORKSCREW FT |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
| Date of Event |
06/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient suffered from an osteolysis after the usage of an arthrex biocomposite screw.No further information received.*** update dw 26-feb-2024: further information was provided that the first arthrex employee was made aware of this issue on 22nd june 2023.The initial surgery was performed on (b)(6) 2020 with an ar-2323bcc and ar-1927bcf.The patient returned due to pain for a mri in (b)(6) 2023.*** update dw 28-feb-2024: further information were provided that no further treatment was performed due to this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint is confirmed based on the mri photo that the customer provided, which displays osteolysis around the implant.The most likely cause for the reported failure can be attributed to a patient-specific event of the immune response allograft tissue and/or a previous history of infection or arthritis of the patient that was not reported due to the patient developing osteolysis.
|
| |
|