• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SYNERGY RESECTION SHAVER CONSOLE; POWERED SURG ORTHOPEDIC INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SYNERGY RESECTION SHAVER CONSOLE; POWERED SURG ORTHOPEDIC INSTR Back to Search Results
Model Number SYNERGY RESECTION SHAVER CONSOLE
Medical Device Problem Code Device Emits Odor (1425)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during shoulder surgery a "popping" sound was heard, a burning smell was noticed and the device shut off.It was further reported that the event occurred after 9 minutes.The device itself was unpacked and connected prior to this surgery.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY RESECTION SHAVER CONSOLE
Common Device Name
POWERED SURG ORTHOPEDIC INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18856453
Report Number1220246-2024-01337
Device Sequence Number8239639
Product Code HWE
UDI-Device Identifier00888867123854
UDI-Public00888867123854
Combination Product (Y/N)N
Initial Reporter CountryIC
PMA/510(K) Number
K932699
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSYNERGY RESECTION SHAVER CONSOLE
Device Catalogue NumberAR-8305
Device Lot NumberEAH230303
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2024
Initial Report FDA Received Date03/07/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-