| Model Number |
EVOLUTR-23-US |
| Medical Device Problem Codes |
Perivalvular Leak (1457); Insufficient Device Problem Information (3190); Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Codes |
Valvular Insufficiency/ Regurgitation (4449); Aortic Valve Insufficiency/ Regurgitation (4450); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that 5 years 1 month following the implant of this transcatheter bioprosthetic valve, the valve failed.The failure mode is unknown.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
Additional information was received that 4 years and 10 months following the valve implant, patient became symptomatic and diagnostic transesophageal echocardiogram (tee) was performed.5 years following the valve implant diagnostic tee was performed again and the mode of valve failure was moderate paravalvular leak (pvl) with moderate to severe aortic regurgitation.The patient was referred to a different hospital for possible intervention.The current patient health status is stable but symptomatic.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional codes corrected data: e3- occupation from hcp to non-hcp h2 - report source from hcp to consumer medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|