• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER Back to Search Results
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 02/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the sterilizer and found the unit to be operating properly.No issues with the function or operation of the sterilizer were identified and the sterilizer was returned to service.The reported burn is attributed to the employee not wearing proper ppe, specifically gloves, while operating their v-pro max sterilizer.The v-pro max sterilizer operator manual (1-2) states, "danger - chemical injury hazard: any visible liquids in the chamber or on the load must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.When handling hydrogen peroxide, wear appropriate personal protective equipment." the v-pro max operator manual (1-2) states, "ansi/aami st58, 2013, recommends using chemical-resistant gloves when using the sterilization unit." additionally, user facility personnel should ensure all instruments are properly dry prior to placement in the v-pro max sterilizer.The v-pro max operator manual, (a-1) states, "dry all items thoroughly.Ensure all moisture is removed from all internal parts (including lumens).If not, residual hydrogen peroxide may remain at cycle completion and/or a cycle abort occurs.Only dry items are to be placed in sterilization unit." the technician counseled user facility personnel on the proper use and operation of their v-pro max sterilizer, specifically the importance of wearing proper ppe and properly drying instruments.No additional issues have been reported.
 
Event or Problem Description
The user facility reported that an employee experienced a burn on their hands while handling items that were processed in a v-pro max sterilizer.The employee was taken to the emergency room and received medical treatment (bandage).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO MAX STERILIZER
Common Device Name
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
MDR Report Key18856945
Report Number3005899764-2024-00009
Device Sequence Number12832777
Product Code FLE
UDI-Device Identifier00724995000028
UDI-Public00724995000028
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2024
Initial Report FDA Received Date03/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-