HOLOGIC, INC MYOSURE XL; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 50-501XL
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) a myosure procedure was performed and during the procedure, the doctor was resecting a calcified fibroid and had used 2 xl devices that had both become dull from cutting and was opening a 3rd device to continue removal.The doctor plugged in the 3rd device and began to continue the resection.The doctor was applying some pressure/torque to the device to keep cutting the fibroid.There was an audible change in sound, then the vibration stopped, then the machine stopped.During resection with the 3rd device, the screen on the myosure control unit went black and the device stopped.The doctor removed the device and staff power cycled the control unit to continue the procedure.The doctor reintroduced the scope after the power cycle and visualized metal shavings in the cavity.The doctor then opened a myosure reach to try and remove the metal shavings from the cavity.It was removed as many shavings as possible and the doctor chose to abort the procedure as they were approaching the deficit limit.At this time, the patient is not being kept for further observation and was not provided with any additional medication or treatment.No additional information available.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.3 devices were involved in this case: 1222780-2024-00094 and 1222780-2024-00095.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE XL
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough 01752
Manufacturer (Section G)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
MDR Report Key18857070
Report Number1222780-2024-00093
Device Sequence Number3313946
Product Code HIH
UDI-Device Identifier15420045505094
UDI-Public(01)15420045505094(17)260906(10)23J20RL
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K142029
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/06/2026
Device Model Number50-501XL
Device Catalogue Number50-501XL
Device Lot Number23J20RL
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/29/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/07/2024
Supplement Report FDA Received Date05/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexFemale
-
-