| Model Number |
PX12N |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
02/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, when taking a blood sample in a neo natal patient with this disposable pressure transducer connected to an arterial line, lot of air came into the syringe and reasonable but unknown blood quantity leaked.Then, it was confirmed there were no loose connections and when flushing the system a leakage was found through the stopcock.Patient demographic unable to be obtained.No further information is available.There was no allegation of patient injury.The device was available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
| |
|
Additional Manufacturer Narrative
|
|
The device involved in this complaint was not available for evaluation.Without return of the product, edwards is unable to perform a complete investigation of the reported event.It is not possible to determine what factors may have contributed to it, and therefore no actions could be planned.Since the affected unit was not returned for evaluation, a product non-conformance or device failure could not be confirmed.As such, based on available information, a definite root cause is unable to be determined at this time.The manufacturing records were reviewed and there is no indication of a related nonconformance; all process parameters were met and inspections passed successfully.Additionally, there are manufacturing controls as leak test, manuafacturing continuous monitoring sampling or qa sampling inspection to avoid potential causes related.
|
| |
|
Search Alerts/Recalls
|