| Catalog Number |
230774001 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It looks like the tip of rod for reaming guide holder has snapped.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Added: h6: (impact code).If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received.A.Was there a surgical delay? if yes, what is the duration of the delay? no delay, more than a minute or so.B.Was/were there any adverse consequence/s that affected the patient, because of the reported event? none reported.C.Did this event occur intra op? yes.This happened, during the operation.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint(b)(4).Investigation summary rod for reaming guide holder, 230774001 holder for reaming guides.It looks like the tip of rod for reaming guide holder has snapped.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the device found that holder for reaming guides had light scratches on the overall surface and presents a used appearance.The observed condition does not impacts the device functionality.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test for the rod for reaming guide holder was performed by assembling into the holder for reaming guides, the device could be assembled successfully.The overall complaint was unconfirmed as the observed condition of the holder for reaming guides would not contribute to the complained device issue.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot :as no defect was found, a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.H10 additional narrative: corrected: h3.
|
| |
|