Catalog Number 113643026
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Unspecified Tissue Injury (4559)
Date of Event 02/28/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Revision of total shoulder replacement - converted to reverse shoulder due to rotator cuff failure causing pain and reduced function.Right shoulder.All components explanted and replaced with reverse shoulder.Doi: (b)(6) 2022; doe: (b)(6) 2024.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : revision of total shoulder replacement - converted to reverse shoulder due to rotator cuff failure causing pain and reduced function.Right shoulder all components explanted and replaced with reverse shoulder.The product was not returned to depuy synthes, however photos were provided for review.The x-ray investigation revealed nothing indicative of a device nonconformance.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the crosslink anchor pg glenoid 52 would not contribute to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device 113643026/jg3091 and no non-conformances were identified.Device history review : a manufacturing record evaluation was performed for the finished device 113643026/jg3091 and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSLINK ANCHOR PG GLENOID 52
Common Device Name
GLENOID (STEPTECH AND ANCHOR PEG) : SHOULDER GLENOID
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18857365
Report Number1818910-2024-05344
Device Sequence Number3325771
Product Code KWS
UDI-Device Identifier10603295007012
UDI-Public10603295007012
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K052472
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue Number113643026
Device Lot NumberJG3091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer05/28/2024
Initial Report FDA Received Date03/07/2024
Supplement Report FDA Received Date05/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLBL UNITE ANT BODY 135 SZ 14; GLOBAL UNITE HEAD 52X15 ECC; GLOBAL UNITE STD STEM SZ 14
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-