• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 GLOBAL UNITE HEAD 52X15 ECC; GLOBAL UNITE IMPLANTS : HUMERAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 GLOBAL UNITE HEAD 52X15 ECC; GLOBAL UNITE IMPLANTS : HUMERAL TRIALS Back to Search Results
Catalog Number 110052600
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Unspecified Tissue Injury (4559)
Date of Event 02/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Revision of total shoulder replacement - converted to reverse shoulder due to rotator cuff failure causing pain and reduced function.Right shoulder.All components explanted and replaced with reverse shoulder.Doi: (b)(6) 2022.Doe: (b)(6) 2024.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary revision of total shoulder replacement - converted to reverse shoulder due to rotator cuff failure causing pain and reduced function.Right shoulder all components explanted and replaced with reverse shoulder ¿the product was not returned to depuy synthes, however photos were provided for review.See attachment [img_9504.Jpeg].The x-ray investigation revealed nothing indicative of a device nonconformance.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the global unite head 52x15 ecc would not contribute to the reported adverse event.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing record evaluation was performed for the finished device 110052600/527776 and no non-conformances were identified.Device history review a manufacturing record evaluation was performed for the finished device 110052600/527776 and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLOBAL UNITE HEAD 52X15 ECC
Common Device Name
GLOBAL UNITE IMPLANTS : HUMERAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18857366
Report Number1818910-2024-05347
Device Sequence Number3325772
Product Code HSD
UDI-Device Identifier10603295004721
UDI-Public10603295004721
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K170748
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number110052600
Device Lot Number527776
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer05/28/2024
Initial Report FDA Received Date03/07/2024
Supplement Report FDA Received Date05/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/27/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CROSSLINK ANCHOR PG GLENOID 52; GLBL UNITE ANT BODY 135 SZ 14; GLOBAL UNITE STD STEM SZ 14
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-