| Model Number |
HD SCORPION NEEDLE (5 PACK) |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 02/21/2024, it was reported by a facility representative via email that an ar-13999hdn scorpion needle tip broke off.The piece remains inside the patient.This occurred during use in a case.Doctor elected to bypass x-ray confirmation of needle tip placement in patient as well as any surgical intervention as it would be more traumatic to attempt to remove it.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Visual inspection revealed that the needle of the hd scorpion needle had broken off and bent.Functional testing was not conducted due to the damage to the tip.The most likely cause of the reported failure is a use error.Per dfu-03920-sub-eo, to help avoid needle breakage and potential patient injury: do not re-sterilize or reuse the scorpion¿ suture passer needle.Avoid striking bone or using excessive force to pass the needle through fibrous/calcific tissue.If excessive force is encountered, replace the needle, reposition the device, and pass in a different area.Ensure a secure grasp of tissue to prevent the needle from buckling under the tissue.Avoid dry, repetitive actuations outside the surgical site.Abrasion of the needle surface can occur that may inhibit proper function.
|
| |
|
Search Alerts/Recalls
|