| Model Number |
ACL TIGHTROPE RT W/ DEPLOYING SUTURE |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the white suture of the device tore completely during insertion of the device.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint allegation is confirmed.One unpackaged ar-1588rt-j acl tightrope® rt was received for investigation.Visual evaluation noted the white loop suture was damaged - it was broken and the tails were frayed.The button was visually evaluated under a magnifier and no sharp edges were observed.Functional testing was unable to be performed due to the damage to the device.The most likely cause for the reported failure can be attributed to user error due to excessive force used during suture tensioning.Refer to investigation photos.
|
| |
|