• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD VIO 36IN 1 S/A CT; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD VIO 36IN 1 S/A CT; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number J359H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that patient underwent a gynecological procedure on (b)(6) 2024, and suture was used.During the operation, when the gynecologist had started to suture, the needle pulled off.Further information has been requested.
 
Additional Manufacturer Narrative
(b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Was there any adverse consequence associated with the patient? please provide the product code related to the vicryl involved in this event to date it has been reported that the device will not be returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002 ¿ device not returned a review of the batch manufacturing records was conducted, and no related non-conformances were identified.Related events captured via 2210968-2024-02679.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Date sent to the fda: 3/22/2024.Additional information was requested, the following was obtained: - was there any adverse consequence associated with the patient? - no patient consequence.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 36IN 1 S/A CT
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key18861180
Report Number2210968-2024-02680
Device Sequence Number3314578
Product Code GAM
UDI-Device Identifier10705031039278
UDI-Public10705031039278
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberJ359H
Device Lot NumberTPBCSLD0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2024
Supplement Date Received by Manufacturer03/11/2024
Initial Report FDA Received Date03/07/2024
Supplement Report FDA Received Date03/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-