• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EPOCAL INC. EPOC BLOOD ANALYSIS SYSTEM; EPOC READER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EPOCAL INC. EPOC BLOOD ANALYSIS SYSTEM; EPOC READER Back to Search Results
Catalog Number 11413475
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the information provided by the customer, the cause of this event is user error.No further investigation required.
 
Event or Problem Description
The customer stated that an operator logged into their epoc instrument to run a test on (b)(6) 2024 at 06:50.Patient 1 was scanned, but the user received an error code and was unable to run the test at that time.Another user then used the same epoc instrument to run a test at 12:55 on patient 2, and the original pid for patient 1 was still programmed into the host at the time of testing so the results for patient 2 were sent to the chart of patient 1.The results were questioned by the physician and blood for patient 2 was needed to be recollected for retesting.The ventilator settings for patient 1 were changed due to the incorrect results and were changed back after retesting a 15:47.There is no report of harm due to the change in treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPOC BLOOD ANALYSIS SYSTEM
Common Device Name
EPOC READER
Manufacturer (Section D)
EPOCAL INC.
2060 walkley road
ottawa, ontario K1G 3 P5
CA  K1G 3P5
Manufacturer (Section G)
EPOCAL INC.
2060 walkley road
ottawa, ontario K1G 3 P5
CA   K1G 3P5
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
MDR Report Key18861413
Report Number3002637618-2024-00016
Device Sequence Number19661639
Product Code CHL
UDI-Device Identifier00630414606095
UDI-Public630414606095
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K200107
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number11413475
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2024
Initial Report FDA Received Date03/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-