| Catalog Number |
11413475 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the information provided by the customer, the cause of this event is user error.No further investigation required.
|
| |
|
Event or Problem Description
|
|
The customer stated that an operator logged into their epoc instrument to run a test on (b)(6) 2024 at 06:50.Patient 1 was scanned, but the user received an error code and was unable to run the test at that time.Another user then used the same epoc instrument to run a test at 12:55 on patient 2, and the original pid for patient 1 was still programmed into the host at the time of testing so the results for patient 2 were sent to the chart of patient 1.The results were questioned by the physician and blood for patient 2 was needed to be recollected for retesting.The ventilator settings for patient 1 were changed due to the incorrect results and were changed back after retesting a 15:47.There is no report of harm due to the change in treatment.
|
| |
|