EXACTECH, INC. SPECIFIC DEVICE NOT REPORTED; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
300-01-13 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Unspecified Infection (1930); Inflammation (1932); Implant Pain (4561)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(d10) concomitant device(s): 320-10-10 - equinoxe reverse tray adapter plate tray +10: 5232147.320-02-42 - rs expanded glenosphere 42mm, +4mm offset: 5645362.320-15-05 - eq rev locking screw: 5600768.320-20-00 - eq reverse torque defining screw kit: 5667604.(h3) pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported by the equinoxe shoulder study, the patient had a left tsa on (b)(6) 2019 and presented with deep infection, on (b)(6) 2022.Patient was seen on (b)(6) 2022.Describes significant pain shoulder while attempting to lift a 50lbs.Bag 6 weeks prior to visit.Work up w/ ct indicated rct.Revision was recommended to convert to rtsa.Patient delayed surgery for over a year.Intra-op findings showed an intact rc.Significant synovitis was encountered, and a deep infection was suspected.A stage revision was carried out by placing a spacer.The outcome of this event is considered resolved on (b)(6) 2024.The action taken was revision on with the removal of the standard humeral stem, replicator plate, torque screw, humeral head, glenoid.The case report form indicates that this event is possibly related to the device and/or to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the sterile certificates and the final endotoxin test reports was performed.All lots were accepted to the requirements.The reason for the reported revision due to infection cannot be conclusively determined; however, it may be related to an underlying patient condition.B3: corrected.D1: corrected.D4/d6: device, serial #, udi #, expiration, implant and explant dates unknown.G3: manufactured dates unknown.G4: pma 510k cannot be determined.H6: corrected health effect, component, and investigation clinical codes.
|
| |
|