| Catalog Number |
100/133/060 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 - other: device has not been returned to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon was punctured even before being used.The malfunction was identified in the pre-use test before placement by the doctor.There is a sample available for evaluation.There was no patient involvement, harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D9: date returned to mfg.: 4/30/2024.H3 and h6.Evaluation codes: updated.Two (2) device samples were received inside the original package; the package was unsealed.Broken cuff was observed in both samples.The complaint was confirmed.The root cause was unknown.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.
|
| |
|