| Model Number |
10FCC20 |
| Medical Device Problem Codes |
Leak/Splash (1354); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ablation procedure, microbubbles were observed in the sidearm of the sheath.The bubbles seemed to be more pronounced during aspiration of the sheath upon catheter removal.Despite the issue, the case was completed successfully. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: one image file and the 10fcc20 sheath with lot number 0012109466 was returned and analyzed.The image showed microbubbles were observed in the side port tube of the sheath.Visual inspection of the handle area was performed.The inspection identified a kink at the side port tube.The tip of the sheath was intact with no anomaly.The steering mechanism and deflection test were performed.No anomaly was discovered.The lab test dilator was inserted into the sheath and retracted several times, without any friction.The lab test dilator was snap-locked to the sheath and was tight.Visual inspection and magnifying with a high-resolution microscope showed the valve housing was intact without any issue.The performance test with the sentinel blackbelt leak tester was performed and the pressure test with 45 pounds per square inch gauge (psig) showed the pressure decay in the device was 0.060 psig and in an acceptable range and the vacuum test with a negative pressure of 8.5 psig showed the pressure decay in the device was 0.008 psig and in an acceptable range.In conclusion, the reported issue of air ingress was not observed/reproduced with the returned sheath.The sheath failed the returned product inspection due to a kink observed on the side port tube.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|