Model Number PVF-XL
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Atrial Fibrillation (1729); Heart Block (4444)
Date of Event 02/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3 other text : device remains implanted.
 
Event or Problem Description
The manufacturer was informed of the following event through mantra study database.Reportedly, a perceval plus valve size xl was implanted in the patient on (b)(6) 2024.Based on information received, patient had chb with afib on (b)(6) 2024.The patient needed a permanent pacemaker which was implanted on (b)(6) 2024 to resolve the issue.As reported, the patient's clinical history included coronary artery disease; arrhythmia: afib; systemic hypertension; diabetes mellitus: type ii, insulin dependent; dyslipidemia; renal insufficiency / renal disorder (dialysis ongoing); percutaneous coronary intervention (stenting 2019); and was in nyha class iii.Based on the information available, prior and concomitant anticoagulant/antiplatelet medications are warfarin, eliquis and aspirin.
 
Additional Manufacturer Narrative
The manufacturing and material records for the perceval plus heart valve and stent, model # pvf-xl-us, s/n # (b)(6), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a (model # pvf-xl-us) perceval plus heart valve at the time of manufacture and release.Based on the information available, the definitive root cause of the reported event cannot be established.However, from the document review performed, no manufacturing deficiencies were noted.Conduction disorders are common in patients with aortic valve diseases.As a result, it is common for patients to receive permanent pacemaker implantation after aortic valve replacement.Risk factors include preexisting conducting disease and preoperative aortic regurgitation.This event is, therefore, a known, inherent risk of the procedure, and it is listed among the potential adverse events in the perceval plus ifu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEVAL PLUS SUTRUELESS AORTIC HEART VALVE
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer (Section G)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer Contact
laura mannino
5005 north fraser way
burnaby, bc 
MDR Report Key18861790
Report Number3004478276-2024-00120
Device Sequence Number8282570
Product Code LWR
UDI-Device Identifier00896208000825
UDI-Public(01)00896208000825(240)PVF-XL(17)270421
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P150011/S013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPVF-XL
Device Catalogue NumberPVF-XL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2024
Supplement Date Received by Manufacturer05/31/2024
Initial Report FDA Received Date03/07/2024
Supplement Report FDA Received Date06/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
Patient Weight114 KG
-
-