| Model Number |
61024BX |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a procedure to access the ureter, the surgeon had difficulty inserting the sheath.When the sheath was taken out of the patient, there was a crack noted near the tip and the tip subsequently broke when placed on the operating table.The procedure was completed with a similar device.There was an unspecified delay; however, there were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.New information added to the following fields: h6 and h11.Corrected field: h6 - type of investigation (code 4111 is not applicable to this event).The device history record was unable to be reviewed since the exact manufacture date is unknown.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The device was not returned to olympus for inspection, therefore the customer's reportable malfunction could not be confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, is likely that the tip broke occurred due to dilator performance impact caused by: aging of the dilators throughout the lifecycle of the product, and ultraviolet (uv) exposure.Three attempts were made to obtain additional information regarding the reported event, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|