• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM JET7 REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM JET7 REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 5MAXJET7
Medical Device Problem Codes Fracture (1260); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the right internal carotid artery (ica) using a penumbra system jet7 reperfusion catheter (jet7), neuron max 6f 088 long sheath (neuron max), and a guidewire.During the procedure, the physician noticed that the proximal end of the jet7 was fractured and leaking after completing the first pass.Therefore, the jet7 was removed.The procedure was completed using another aspiration catheter and the same neuron max.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned jet7 confirmed a leak coming from underneath the strain relief.If the device is manipulated against resistance or at an angle, damage such as a kink and a subsequent fracture near the hub may occur.Further evaluation revealed kinks on the distal shaft, and the distal tip was shaped.The shaped distal tip may have contributed to resistance.The kinks on the distal shaft may have also occurred during advancement against resistance during the procedure.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM JET7 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key18862338
Report Number3005168196-2024-00075
Device Sequence Number3307157
Product Code NRY
UDI-Device Identifier00814548019857
UDI-Public814548019857
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K173761
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2024
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2024
Device Catalogue Number5MAXJET7
Device Lot NumberC31494
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2024
Supplement Date Received by Manufacturer03/07/2024
Initial Report FDA Received Date03/07/2024
Supplement Report FDA Received Date04/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-