| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the treatment couch coordinates are not captured in mosaiq and are not matching the coordinates that the therapists treated the patients with.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Section b3 updated.Section h6 updated.Section h11 updated.The investigation was completed by conducting a thorough evaluation of the product and the reported information.According to the beam records and collected information, mosaiq correctly recorded the couch values for the fields.Truebeam sent the incorrect values to mosaiq due to misconfiguration in the truebeam admin.Mosaiq recorded the data that was sent from the machine.The customer has corrected the misconfiguration for the truebeam machine.Mosaiq did not have any malfunction and is working as designed and intended.Based on the available information, there was no patient mistreatment.Please note that due to an administrative error, the product was initially reported under mosaiq 2.64.210.0001 but this is incorrect and should have been reported under mosaiq 2.83.053 therefore the mfr report number reported against the product is incorrect.A combined report (initial and final) has been submitted under the correct mfr number for this incident capturing all the correct product details (see report 3015232217-2024-0023).
|
| |
|
Search Alerts/Recalls
|