Catalog Number 8065751012
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 02/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported that the small hole was made when the posterior capsule was touched with ophthalmic irrigation aspiration (ia) tip during cortical aspiration.The ophthalmic tip was with scratches or burr.The surgery was completed without product replacement.The procedure type is unknown.The patient had additional anterior vitrectomy surgery.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in sections d.9.H.3.H.6.And h.11.Specific product identifiers (lot number) were not provided and could not be determined at this time.However, all device history records are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this lot number cannot be performed as the lot number is unknown.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.However, a tip was returned for testing on this investigation.The tip was received for testing on this investigation.A visual assessment of the returned sample found several burrs, one located on the shaft and the other on the port.The reported event was confirmed.However, how or when the nonconformities appeared are unknown and the root cause of the reported event is inconclusive.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTREPID AUTOSERT IOL HANDPIECE, I/A TIP
Common Device Name
APPARATUS, CAUTERY, RADIOFREQUENCY, AC-POWERED
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18864432
Report Number2028159-2024-00392
Device Sequence Number20903625
Product Code HQC
UDI-Device Identifier00380657510122
UDI-Public00380657510122
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112425
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065751012
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/29/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2024
Supplement Date Received by Manufacturer05/10/2024
Initial Report FDA Received Date03/08/2024
Supplement Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age29 YR
Patient SexMale
-
-