OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 6,4/7,8 FR. X 430 MM, 7°, STRAIGHT OCULAR ANGLE, 4,2 FR. CHANNEL
|
Back to Search Results |
|
| Model Number |
WA29048A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, the ureteroscope, 6,4/7,8 fr.X 430 mm, 7°, straight ocular angle, 4,2 fr.Channel had a loose insertion tube.The issue occurred during reprocessing.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over nine (9) years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's reportable malfunction was confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the insertion tube is loose occurred due to mechanical force caused by an impact or fall.Olympus will continue to monitor field performance for this device.
|
| |
|