• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. BML HANDLE V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. BML HANDLE V Back to Search Results
Model Number MAJ-441
Medical Device Problem Codes Break (1069); Device Reprocessing Problem (1091)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, the bml handle grip on the operating pipe was weak and came loose.The issue occurred during preperation for a therapeutic lithotripsy procedure.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device was not returned.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.G3 of the initial medwatch is being corrected to feb, 2024 and not 18 feb, 2024 as submitted.Corrections have also been made to b5 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the user did not properly understand the cds (cleaning disinfecting and sterilization) process.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: ·never use excessive force to operate the instrument.This could damage the instrument ·do not operate any parts with excessive force.This may lead to damage to the product.·this instrument is compatible with 2.0 ¿ 3.5% glutaraldehyde solution.However, routine biological monitoring is not feasible with glutaraldehyde and, therefore, it should not be used to sterilize reusable medical devices that are compatible with other methods of sterilization that can be biologically monitored , such as steam sterilization.·although this product is resistant to 2-3.5% glutaral preparations, we recommend autoclaving (high-pressure steam sterilization), which allows you to routinely check the sterilization effect using biological indicators.Place the sealed packages containing the instrument in the autoclave and sterilize in accordance with the conditions listed below.For details on operation of the autoclave, refer to the instruction manual for the autoclave or other manufacturer instructions.After placing the sterilized pack containing this product into the autoclave device, autoclave it under the following conditions.Other conditions should follow various cleaning and sterilization guidelines and manufacturer's instructions.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer's reported event of a weak and loose operating pipe grip was not a reportable malfunction as initially submitted.However, upon further communication, it was found that the customer did not use autoclave sterilization, and instead used their own measures.This mdr is being submitted to note the reportable event of improper sterilization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BML HANDLE V
Common Device Name
HANDLE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18866029
Report Number9614641-2024-00621
Device Sequence Number3306261
Product Code LQC
UDI-Device Identifier04953170218446
UDI-Public04953170218446
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K903529
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-441
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2024
Supplement Date Received by Manufacturer03/21/2024
Initial Report FDA Received Date03/08/2024
Supplement Report FDA Received Date03/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BML-V437QR-30 SINGLE USE MECHANICAL LITHOTRIPTOR V; TJF-Q290V
Patient SexUnknown
-
-