| Catalog Number |
8065751014 |
| Medical Device Problem Codes |
Disconnection (1171); Fitting Problem (2183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that the tube was found disconnected, auto gas fill syringe part was touching a stopper from the beginning, however the syringe was fixed during vitrectomy surgery.The surgery was completed after replacing the product with another product.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information received following submission of the initial report, this event ¿tube found disconnected, syringe part was touching a stopper¿ during surgery does not meet criteria for reporting as a reportable malfunction since it was only issue with auto gas filling not with the cassette.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Based on the information received following submission of the initial report, this event ¿tube found disconnected, syringe part was touching a stopper¿ during surgery does not meet criteria for reporting as a reportable malfunction since it was only issue with auto gas filling not with the cassette.
|
| |
|