• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ETS ARTIC LNR CUTR NO KNIFE; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ETS ARTIC LNR CUTR NO KNIFE; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ATS45NK
Medical Device Problem Codes Failure to Fire (2610); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 3/8/2024.: batch # 891a08.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ats45 device was received with no apparent damage and with a tr45w reload loaded in the device.The reload was received partially fired 1/10 and with the reload lockout spring damaged.The returned device was tested for functionality with test reload and it fired and formed all the staples as intended.The staple line was complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If the re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 891a08, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during a laparoscopic kidney transplantation, the handle did not move at 1st stroke of 1st firing.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS ARTIC LNR CUTR NO KNIFE
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18866307
Report Number3005075853-2024-02048
Device Sequence Number16972524
Product Code GDW
UDI-Device Identifier10705036001089
UDI-Public10705036001089
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberATS45NK
Device Lot Number891A08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2024
Initial Report FDA Received Date03/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TR45W
Patient SexUnknown
-
-