• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL KNIFE KD-655

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL KNIFE KD-655 Back to Search Results
Model Number KD-655Q
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is ongoing, this report will be supplemented when new and relevant information becomes available.
 
Event or Problem Description
Olympus reviewed the following literature titled "v3-1 utility and safety of tech knife in colorectal esd." literature summary endoscopic submucosal dissection (esd) of colorectal tumors is considered a technically difficult and time-consuming procedure.It has been reported that advances in devices contribute to a reduction in the complication rate, and various devices have been developed to date to improve the safety and efficiency of the procedure.Unlike conventional tip-based knives, the tecknife (microtec, inc.) can deliver fluid from the tip of the knife, enabling efficient incision and dissection, which may result in a reduction in esd procedure time.Objective to prospectively evaluate the efficacy and safety of the microtec knife in colorectal esd.Methods (b)(4) colorectal esd cases performed during a 6-month period from (b)(6) 2023 were divided into 2 groups: tk group ((b)(4) cases) using a techknife and dk group ((b)(4) cases) using dualknife j (olympus), and compared.The primary endpoint was dissection speed, and secondary endpoints were esd procedure time, the number of times the local injection needle was inserted and removed, total local injection volume, and intraoperative perforation rate.The authors also performed a subgroup analysis by dividing the patients into two groups: experts with more than (b)(4) esd cases and non-experts with less than (b)(4) esd cases.Results there were no significant differences in patient background (age, gender, comorbidities, anticoagulant medications, localization, gross type, expert / nom-expert) between the two groups.Clinicopathologically, the two groups did not differ in resection specimen diameter (36.5 mm vs 33.6 mm; p=0.22), tumor diameter (30.8 mm vs 27.7 mm; p=0.18), pathologic findings (adenoma/serrated lesion/cancer 36/15/18 vs 27/16/25; p=0.31), and ro resection rate (97.1% vs 94.1%; p=0.12).Although there was no significant difference between the two groups in terms of total localized volume (34.3 ml vs.32.7 mm; p=0.67), the number of device insertions and removals was significantly lower in the tk group than in the dk group (2 vs.3.5 times: p<0.05), resulting in a trend toward faster procedure times in the tk group (42.7 min vs 52.5 min; p=0.13) and significantly faster dissection speed (36.2 mm2 vs 23.4 mm2: p<0.05).Device occlusion was observed in 3 patients (43%) in the tk group, but not in the dk group.Intraoperative perforation occurred in 2 patients (2.9%) in the dk group, but not significantly (p=051).The results of the expert/non-expert group were also compared with those of the tk and dk groups, respectively.Experts in the tk group had significantly fewer device insertions and removals (2 vs.3.6 times; p<0.05), shorter esd procedure times (37.5 min vs 50.3 min; p=0.07), and significantly faster dissection speeds (36.8 mm2 vs 26.5 mm2; p<0.05).On the other hand, there was no difference in any of the parameters between the tk and dk groups in the non-expert group.Conclusion compared to the dualknife j, the techknife may reduce procedure time by reducing the number of times the device is inserted and removed, and by performing dissection more efficiently, suggesting that the use of the techknife may be more effective, especially for experts.Type of adverse events/number of patients intraoperative perforation (n=2).
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial and/or lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the relationship between the device and the adverse event cannot be confirmed.There was no complaint reported on the subject device.There is no evidence of an olympus device malfunction.Therefore, the root cause cannot be determined.This supplemental report includes additional information received from the author.B5 updated accordingly.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
An olympus device did not cause or contribute to the adverse events described in the article.Also, it was confirmed that an olympus device malfunction did not occur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL KNIFE KD-655
Common Device Name
SINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18868702
Report Number9614641-2024-00627
Device Sequence Number15437752
Product Code KNS
UDI-Device Identifier04953170405631
UDI-Public04953170405631
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K171158
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberKD-655Q
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2024
Supplement Date Received by Manufacturer03/13/2024
Initial Report FDA Received Date03/08/2024
Supplement Report FDA Received Date03/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-