| Model Number |
CI-1400-01 |
| Medical Device Problem Codes |
Loose or Intermittent Connection (1371); Mechanical Problem (1384); Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The recipient is reportedly experiencing intermittent lock, despite device testing within normal limits.External equipment was exchanged, however, the issue did not resolve.A ct scan revealed no abnormalities with the implant proper lead placement.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.Connection was reportedly obtained using a new processor, with the magnet strength adjusted and the recipient's issues have resolved.The recipient has resumed device use.No further details will be provided.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: section d.6b.Additional information: section d.9.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The device passed the external visual inspection.The photographic imaging inspection revealed broken antenna coil wires.System lock was obtained intermittently.The device passed the electrical and mechanical tests performed.This device had broken antenna coil wires.A corrective action was implemented.This version of the hires 90k device is no longer distributed.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|