| Catalog Number |
72201518 |
| Medical Device Problem Codes |
Break (1069); Dent in Material (2526); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, a white material of the tip of the abrader torn, it was part of the inner blade.And it also broke on the upper part of the blade.The procedure was completed without delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the reported device was received for evaluation.A visual inspection revealed that it is not in its original packaging.The white polyester shrink tubing on the inner shaft is cut and scuffed.The inner shaft is fractured toward the distal end.The outer shaft is dented at the cutting window and the interior of the hub is damaged at the insertion point for the inner shaft.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include an application of unintended inappropriate or excessive force to the device.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|