| Catalog Number |
110027734 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Osteopenia/ Osteoporosis (2651)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer has not indicated whether the product will be zimmer biomet for investigation, and it may remain implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery, the surgeon questioned why the central screw was not reading the amount that was recorded on the screw schematic.It was observed that the implant most likely was not seated correctly.After a review of the 3d-modeled surgical plan, it was discovered that the implant had been implanted upside down.No patient consequences or plans for revision have been communicated at this time.Attempts have been made, and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.D10: associated product information, part number (lot number): ¿ 110031378 (66518780).¿ 110019066 (66518781).¿ 110019066-01 (66518782).¿ 110019066-02 (66518783).If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4; b5; d2; g2; g3; g6; h1; h2; h3; h6.The following sections were corrected: a2; b6.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: single ap view of the right shoulder demonstrates a reverse total shoulder arthroplasty with inferior positioning of the glenoid component with respect to the scapula.Central screw appears to be protruding through the scapula.No dislocation.Osteopenia is present.Findings suggestive of malposition of the glenoid component of a right reverse total shoulder arthroplasty.The root cause of the reported issue is attributed to user error as the glenoid was implanted upside down.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|