| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/10/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: report source france the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a lag screw reamer for cmn nail broke during primary surgery.A small piece was left and is still inside the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4): this follow-up report is being submitted to relay additional information.The following sections were updated.The lag screw reamer was returned for investigation.The shaft of the instrument shows signs of normal usage, but it is overall inconspicuous.The cutting edge of the instrument is fractured.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Review of the complaint history identified additional similar complaints for the reported item and the part and lot combination.Based on the analysis of the returned instrument, a fracture of the lag screw reamer can be confirmed.The review of the surgical technique suggests that using the wrong ccd angle during assembly of the lag screw reamer on the targeting guide could potentially lead to impingement of the tip of the lag screw on the nail, which might contribute to the fracture of the reamer.However, with the available information, it is impossible to reproduce the reported event and a definitive root cause cannot be established.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|