MAUDE Adverse Event Report: ZIMMER GMBH ZNN¿¢, CMN LAG SCREW REAMER, SHORT; TRAUMA, INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH ZNN¿¢, CMN LAG SCREW REAMER, SHORT; TRAUMA, INSTRUMENTS Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/10/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: report source france the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that a lag screw reamer for cmn nail broke during primary surgery.A small piece was left and is still inside the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4): this follow-up report is being submitted to relay additional information.The following sections were updated.The lag screw reamer was returned for investigation.The shaft of the instrument shows signs of normal usage, but it is overall inconspicuous.The cutting edge of the instrument is fractured.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Review of the complaint history identified additional similar complaints for the reported item and the part and lot combination.Based on the analysis of the returned instrument, a fracture of the lag screw reamer can be confirmed.The review of the surgical technique suggests that using the wrong ccd angle during assembly of the lag screw reamer on the targeting guide could potentially lead to impingement of the tip of the lag screw on the nail, which might contribute to the fracture of the reamer.However, with the available information, it is impossible to reproduce the reported event and a definitive root cause cannot be established.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZNN¿¢, CMN LAG SCREW REAMER, SHORT
Common Device Name
TRAUMA, INSTRUMENTS
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18873645
Report Number0009613350-2024-00090
Device Sequence Number2957916
Product Code HTO
UDI-Device Identifier00889024276260
UDI-Public(01)00889024276260(11)230614(10)4504770497
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K091566
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00249000364
Device Lot Number4504770497
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2024
Supplement Date Received by Manufacturer07/10/2024
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date07/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-